Medical equipment, bedside monitor, invasive patient monitor, ICU monitor, measure artery monitor
medical monitor, hospital monitor , invasive monitor
MONITOR (for measure artery, invasive)
ICU Monitor
Multi-Para Patient monitor
bedside monitor
ICU Monitor
12.1 "Standard 6 Parameters
ECG, NIBP,SpO2,RESP, TEMP,HR/PR
Features &Benefits:
- 12.1" color TFT display, high resolution: 800×600dpi;
- Separated para board: ECG, NIBP, SpO2 board;
- Full touch-screen enables intuitive operation by clicking one the specific parameters or the waveform in real time, standard configuration includes the navigation knob.
- Real time S-T segment analysis, and ARR analysis;
- SpO2 Pitch Tone Variation and Drug Dose Calculation;
- Parameter waveform and characters color selectable
- Multi-display: Standard, Large font, Trend Coexis, OxyCRG dynamic
- Standby mode for Energy saving and suspend monitoring;
- 7-lead ECG waveforms display in phase;
- Capture dynamic waveforms;
- Built-in rechargeable lithium battery, and battery volume checking,
- AC power checking
- Efficient resistance to interference of defibrillator and HF knife;
- SINNOR F-6 SpO2 technology, VS NELLCOR accuracy;
- Networking capacity and TCP/IP networking platform enable abundant future upgrade
- Time delay switch-off function keep the monitor data restart monitor;
- Complete and flexible mounting solution for transportation and in-patient monitoring
Standard parameters: 5-lead ECG, SpO2(F-5),NIBP, TEMP,RESP,PR,HR.
Optional: Touch Screen, Thermal Recorder, 3/12-lead ECG, 2/4-IBP,4-TEMP,2-SpO2, Nellcor/Masimo SpO2,BIS, Phaisein/ Respironics Multi-gas, ICG/CO, FHR, FM, TOCO,
Performance Specifications
Display: 12.1"color TFT
Rolling and refreshing waveform display
Resolution: 800×600
Multi display selectable including:
Large-font display
Trend coexist display
OxyCRG dynamic view display.
Bed-to-bed view display
Trace: 9 waveforms(7 ECG, 1 SPO2 and 1 RESP)
Sweep speed:12.5mm/s,25mm/s,50mm/s
Indicator: Power/battery indicator light
QRS beep and alarm sound
Battery: Rechargeable lead acid cell, 12v/4AH
Max.24 hour for charging, 4 hours for continue working
Trend: Parameter graphic and tabular trends:
5s/piece,8 hours;
1min/piece,168 hours(24hour×7days)
5min/piece,1000 hours.
Storage: NIBP: 1000 groups
Alarm: 200 groups
Full disclosure waveforms: 3600S
SD card external storage
Alarm: User-adjustable High, Medium and Low limits 3-level
Audible and visual alarm
Networking: Connected to central monitoring system
TCP/IP netting platform
Recorder: Build-in, thermal array
Plethysmogram caveform:2 channels
Record mode: manual, on alarm, time-defined
Recording width: 50mm
Printing speed: 50mm/s
Recording type: Frozen waveform record
NIBP recall record
Trend table record
Alarm record
Fixed -time record
Standard parameter:
ECG:
Lead mode: 5 -lead(R,L,F,N,C)
Lead selection : I,II,III,avR,avL,avF,V
Waveform: 3 and 7 channel selectable
Gain selection: 0.5mm/mv,1mm/mv,2mm/mv
Sweep speed: 12.5mm/s;25mm/s;50mm/s
Heart rate range: Adult: 15~300bpm;
Neonate/pediatric:15~350bpm
Accuracy: +1bpm or +1%,whichecer is greater
Resolution: 1bpm
Filter: surgery mode:1~20Hz
monitor model:0.5~40Hz
Diagnostic mode:0.05~130Hz
Scaling signal: 1mv,+3%
Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
Alarm range: 15~350bpm
S-T segment detecton:
Measurement range: 2.0mV~+2.0mV
Alarm Range: -2.0mV~ +2.0mV
Accuracy: -0.8mV ~+0.8Mv
Error: +0.02Mv
Arrhythmia analysis: YES
SPO2
Measurement range: 0~100%
Resolution: 1%
Accuracy: +2%(70-100%);0-69% unspecified
Alarm range 0~100%
Pluse Rate:
range: 20~300bpm
Resolution: 1bpm
Error: +1bpm or +2%,whichever is greater
NIBP
Method: Digital Automatic oscillometric
Operation mode : Manual/Automatic/continuous
Auto measurement time: Adjustable(1~480min)
Measurement Unit: mmHg/Kpa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement tange:
Range of Systolic pressure: Adult: 40~270mmHg
Pediatric: 40~220mmHg
Neonate: 40~135mmHg
Range of mean pressure: Adult: 20~235mmHg
Pediatric: 20~165mmHg
Neonate: 20~110mmHg
Range of diastolic pressure: Adult: 10~215mmHg
Pediatric: 10~150mmHg
Neonate :10~100mmHg
Over-pressure protection: Double safety protection
Resolution: 1 mmHg
Alarm: Systolic.Diastolic,Mean
RESPIRATION
Method: Thoracic impedance
Measurement range: Adult:7~120rpm;
Neonate/Pediatric:7~150rpm
Apnea alarm: YES,10~40s
Resolution: 1rpm
Accuracy: +2rpm
TEMPERATURE
Compatible probe: YSI or CYF
Measurement range: 5~50ºC
Resolution : 0.1ºC
Accuracy: +0.1ºC
Refreshing time: about 1
Average measuring time: <10s
FHR
Transducer: Multi-crystal, Pulsed Doppler
Measurement range: 50~210 BPM
Working frequency: 1 MHz
Strength: <5mW/cm2
Signal processing:
special DSP system and modern recognization.
Resolution: 1BPM
Accuracy: ±1BPM
Alarm Range: High: 160,170,180,190 BPM,
Low: 100,110,120 BPM
FM
Manual button marking,
automatic FM identifying function
TOCO Measuring
Transducer: External pressure transducer
Measurement range: 0~100 units
Resolution: 1rpm
Accuracy: ±2 rpm
IBP
Channel: 2 channels
Range: -50-300mmHg
Resolution: 1mmHg
Accuracy: ±4mmHg(±4%)
Unit: mmHg, Kpa
Converter Sensitivity: 5mV/V/mmHg
Transducer sites: ART/PA/CVP/LAP/RAP/ICP
EtCO2( Sidertream CO2)
Measurement range: 0~99mmHg
Accuracy: +2mmHg (0~40mmHg)
Sampling Range: 100ml/min
Sampling rate accuracy: 15%
Respiration rate: 0~120rmp
Respiration accuracy: +2rmp (0~70rmp)
+5rmp (>70rmp)
Respiration time: <240msec (10% to 90%)
Delay time: <2s
EtCO2 (mainstreaam CO2)
Method: Infrared Spectrum
Range: 0.0-10% (0~76%)
Resolution: 1mmHg (0.1%)
Accuracy: <5%(±4.0 mmHg)
Or<10%(of Readings)
Recorder:
Build-in, thermal array
Plethysmogram caveform:2 channels
Record mode: manual, on alarm, time-defined
Recording width: 50mm
Printing speed: 50mm/s
Recording type: Frozen waveform record
NIBP recall record
Trend table record
Alarm record
Fixed -time record
Miscellaneous
Satety:
Safety level: Class I, type CF
Dimendion and Weight
Dimendion: 28*42*48cm
G.Weight: <6.5KS
Operation Environment
Temperature: Working 0~+40ºC
Transportation and storage -20~+60ºC
Humidity: working≤85%
Transportation and storage≤93%
Power: AC 100-240,50/60Hz
Patient range: Neonate,pediatric,and adult patients
Accessories supplies
(1) 5 lead ECG cable
(2) 1 spo2 probes
(3) 1 NIBP prbe
(4) 1 temp probe
(5)1 ground connecting liner
(6) Chest electrode (10pcs/set)