Medical equipment ,Mainstream EtCO2 module
patient monitor, CO2 patient monitor
JYK0015
It is the ideal capnography solution for your monitoring needs. It is mainstream CO2 sensor that plug-in and measurement. There is no need to take up valuable space inside your monitoring system and connects to monitor just with one lemo connector. LH900 features advanced technology for measurement of instantaneous CO2 concentration, respiration rate, end-tidal CO2 concentration, inspired CO2 concentration, inspiration time and expiration time. It is generally used in patient monitor and anesthetic machine to monitor the intubated and non-intubated patients.
Transducer Type: Mainstream CO2 sensor
Principle of Operation: Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
Energy Emitting Device: Proprietary High Efficiency IR Source
Energy Detecting Device: Frequency Stable Thermopile
CO2 Measurement Range: 0% to 20% (0 to 150mmHg)
CO2 Resolution: 0.1 mm Hg
CO2 Accuracy:
±2 mm Hg @ < 5.0% CO2 (at BTPS),
< 10% of reading @ > 5.0%CO2(at BTPS)
Respiration Rate Range: 0 to 150 Breaths Per Minute (BPM)
Respiration Rate Accuracy: ±1 breath
Calibration: No routine user calibration required. An airway adapter. zero is required when changing to a different style of airway adapter.
Data Interface: Highly configurable serial digital interface (TTL/RS232 Level)
Data Output: CO2 gas concentration (mmHg), End-tidal CO2, Inspired CO2, Respiratory Rate.
Input Voltage: 5.00 Volts (± 5%)
Power Consumption: 0.8mW
Warm-up Time: 3 seconds
Response Time: 8mS
Operating Temperature: 5 degree to 50 degree Storage: -40 to 70 degree
Dimensions: 50mm*22mm*36mm
Weight < 95g
Interconnection: Standard Lemo Redel 8-pin plastic Material PSU.
Pin out:
1- VA 5.0V;
2- Shield Shield;
3- DGND Digital return;
4- VA 5.0V;
5- TxD Serial data from the Module;
6- RxD Serial data from Host;
7- AGND Analog return;
8- NC
Shock Impact: IEC TR 6 0721-4-7 Class 7M3 (designed to withstand environments subject to significant
vibrations or high shock levels)
EN6 0068-2-27 Shock
EN60068-2-64 Random vibration
Regulatory:
Designed to meet IEC 6 0601-1-2 , EN55011 CISPIR 11 Class B (Radiated and Conductive Emissions), IEC 6 1000-4-2 Electrostatic Discharge Immunity, IEC 6 1000-4-3 Radiated Immunity,
Designed to comply with 93/42/EEC (MDD CE Marking),
FDA Standards, Minimum Performance and Safety
Requirements for Capnometers and ISO21647.
CO2 Measurement Range |
CO2 Measurement Range |
CO2 Resolution |
0.1 mm Hg |
Respiration Rate Range |
0 to 150 Breaths Per Minute (BPM) |
Respiration Rate Accuracy |
±1 breath |