Medical equipment
Surgical ventilator
TKR-300JII computerized high-frequency first-aid ventilator
(First -aid, easy -to-operate)
It is widely used for clinical salvaging and resuscitate also is a professional device used for respiration salvaging on emergency ambulance.
Feature
A. An opening system is designed to avoid leakage of airway, When ventilating, the intubation may have no casing bag. Little tube may be used for ventilation when resistance of expiration is small, so that mucosa won't be hurt easily.
B. Airway is open during ventilating, suction and ventilation can be done synchronously, so good effectiveness of oxygen combination can be shown without inspiration and expiration synchronizing.
C. Lower airway pressure is good to increase blood discharge and decrease cerebra pressure
D. Suction device, it driving negative pressure is adjustable.
E. Storage battery can be used continuously for four hours after full charged.
F. The system disposes completely with laryngoscope, inlet intubation, catheter of tracheotomy, nasal prongs tube, air tube and so on.
Specifications
Frequency: 10~150 breaths/min
I/E ratio: 1: 1, 1: 1.5~1: 3 and inverse ratio
Driving Pressure: 0~0.15MPa
Negative pressure suction: -0.07~0MPa
CPAP: 0.1~0.8kpa
Volume cycling ventilation: T=10min, I/E=1: 1.5
Adults: T1(100bpm): T2(20bpm)=7: 3 Regular frequency: Tidal volume≥ 400ml/breath Driving pressure: 0.06mpa
Children: T1(100bpm): T2(30bpm)=7: 3 Regular frequency: Tidal volume≥ 300ml/breath Driving pressure: 0.04mpa
Storage battery: 4Ah, 12V; Rechargeable
SIGH: 6 bpm
Alarm: Alarm with sound and light when lack of power, or no air supply.
Maximum Output Oxygen Concentration: 85%
Ventilation Mode: HFJV, CPAP, Volume cycling ventilation; HFJV+SIGH.
Frequency |
10~150 breaths/min |
I/E ratio |
1: 1, 1: 1.5~1: 3 and inverse ratio |
Driving Pressure |
0~0.15MPa |
Negative pressure suction |
-0.07~0MPa |
CPAP |
0.1~0.8kpa |