Customization: | Available |
---|---|
Display: | HD LCD |
Transport Package: | Carton |
Still deciding? Get samples of US$ 220/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Unique Human Voice Alarm system | The device triggered the human voice prompt and stop injection automatically when there is any malfunction, making the infusion process safer and more reliable. |
Rate: | 1ml/h-1,200ml/h |
Flow Rate Accuracy: | Within ±5% (after correct calibration) |
Mechanical precision: | Within ±2% |
Purge Rate: | 100ml/h-1,000ml/h (100 ml/h step) |
Infusion Volume: | 1ml-9999ml |
Total Infusion Volume: | 0.1ml-9999.9ml |
KVO rate: | 1ml/h-5 ml/h (1ml/h step) |
Occlusion: | High: 800mmHg ±200mmHg (106.7kPa±26.7kPa Medium: 500mmHg ±100mmHg(66.7kPa±13.3kPa Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa ) |
Alarms: | End of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out etc. |
Power Source: | AC 100V-240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup |
Bubble detector: | Ultrasonic wave detector; detection sensitivity ≥25μL |
Fuse: | F1AL/250V, 2 pc inside |
Power Consumption: | 30VA |
Operating Condition: | Ambient temperature: +5-+40; Relative humidity: 20-90% Atmospheric pressure: 86.0kpa-106.0kpa |
Transport & Storage Condition: | Ambient temperature: -30-+55 Relative humidity: ≤95% |
Equipment Classification: | Class I, internal power supply, Type CF |